CC13 Ionization Chamber

Standard chamber for clinical use in water phantoms CC13 is the standard chamber for clinical use in water phantoms and for measurements of output factors. The ionization chamber CC13 is intended for absolute and relative dosimetry of photon, electron, and proton beams in radiotherapy. Measurements can be performed in air, water phantoms, or solid phantoms.

 

لینک اطلاعات محصول:

CC13 Ionization Chamber

با ما در تماس باشید
۰۲۱-۲۲۹۱-۶۴۹۴

Standard chamber for clinical use in water phantoms CC13 is the standard chamber for clinical use in water phantoms and for measurements of output factors. The ionization chamber CC13 is intended for absolute and relative dosimetry of photon, electron, and proton beams in radiotherapy. Measurements can be performed in air, water phantoms, or solid phantoms.

لینک اطلاعات محصول:

CC13 Ionization Chamber

Active volume:

0.13cm3

Total active length:

5.8mm

Cylinder length:

2.8mm

Inner diameter of outer electrode:

6mm

Wall thickness:

0.4mm (70 mg/cm2)

Diameter of inner electrode:

1mm

Length of inner electrode:

3.3mm

Polarizing voltage:

±300V (max. ±500 V)

Typical leakage current:

3fA

Typical sensitivity:

3.6nC/Gy

Guard potential:

±300V (max. ±500 V)

Temperature range:

15°C – 35°C

Relative humidity range:

20% – 80%